体细胞核转移(SCNT)/制备杂交瘤转染系统-细胞分析仪器-仪器设备-生物在线
世联博研(香港)科技有限公司
体细胞核转移(SCNT)/制备杂交瘤转染系统

体细胞核转移(SCNT)/制备杂交瘤转染系统

商家询价

产品名称: 体细胞核转移(SCNT)/制备杂交瘤转染系统

英文名称: Somatic-Cell Nuclear Transfer (SCNT)/Hybridoma Production

产品编号: LF101

产品价格: 0

产品产地: 日本

品牌商标: nepagene

更新时间: null

使用范围: null

世联博研(香港)科技有限公司
  • 联系人 :
  • 地址 : 北京市海淀区西三旗上奥世纪中心A座9层906
  • 邮编 :
  • 所在区域 : 北京
  • 电话 : 186****1725 点击查看
  • 传真 : 点击查看
  • 邮箱 : 787852745@qq.com

LF101 Electro Cell Fusion Generator

Somatic-Cell Nuclear Transfer (SCNT)/Hybridoma Production

LF101 Electro Cell Fusion Generator

LF101 Applications

  • Somatic-Cell Nuclear Transfer (SCNT) for Bovine, Pigs, Mice, Rabbits, etc.
  • Hybridoma Production for Monoclonal Antibodies
  • Plant Protoplast Fusion
  • Egg Activation

LF101 Features

point Specialized setting for bovine, pig, mouse and rabbit cloning (somatic cell and embryo nuclear transplant) will result in extremely high fusion probability (for example: more than 90% for bovine cloning).
point Voltage setting with 1V resolution is possible in the specialized setting for mammalian cloning. It is suitable to output precisely the square pulse wave shape and pulse range as preset and accordingly brings about a higher rate of cell fusion probability.
point Post fusion function for damage recovery of the cell can be operated with On/Off switching.
point Specialized setting for plant protoplast and mammalian cell fusion can provide a desirable condition for the cell fusion, which is done with DC pulse (100V, max.) of the precisely controlled voltage and pulse range after producing the pearl chains with DC. It can elevate the rate of hybridoma production (cell fusion) probability.
point Operation of the device is very easy because both of automatic and manual (number of pulses: 1-99) transmitting of the pulse are possible. And the pulse interval can be selected to set between 0.1-9.9 seconds. The device can keep a constant operating condition anytime.
point NEPA GENE provides a variety of electrodes for combination use with the device, which is designed to match well with its various settings through our long accumulated experiences. Combination use of the electrodes with the device enables very easy and efficient gene transfection and cell fusion.

LF101 Specifications

point AC Pulse

Wave Shape Sine Wave
Voltage 0 - 50V (VP-P)
Frequency 0.1 - 3.9MHz Variable
Duration 0 - 99s
Post Fusion 0~9sec (On・Offスイッチ)
Shunt Resistor Greater than 50 ohm

point DC Pulse

Wave Shape Square
Voltage 0 - 999V
0 - 200V (1V resolution)
Pulse Length Range 5~99μs (1μs resolution)
Pulse interval 0.1~9.9s
Number of Pulses 0 - 99 Automatic or Manual
Power 100V (50/60Hz) 2A
Size / Weight 375W×430D×160H mm , 12.7kg